In Adjust, PulStar delivers a series of impulses through the impulse head at the force and mode you choose, and shows how the stiffness changes as you treat. (If your practice uses the term “Mobilize”, the screen shows that instead.)
After the analysis, PulStar suggests vertebrae to treat, starting with the most significant red bars. The number suggested is set by Auto Select Vertebra, and their order by Adjustment Order, in Session preferences. The choice is always yours.
Check the force before every adjustment. The force goes from 5 to 35 in steps of 5. Each spinal region starts at its default force, or at the patient's own force if they have one. A common starting point is 20, just above the analysis force of 15, increasing only if the result you want isn't achieved.
| Mode | How it works |
|---|---|
| Osseous | The default. PulStar sets the rate automatically from the tissue's stiffness, slowing as stiffness drops, and can stop the impulses on its own once the stiffness stops changing. |
| Specific | Holds a rate you choose, from 2 to 35 impulses per second (Hz), with the Rate control. Consider it for larger patients, or when you have a particular rate in mind. |
| Myofascial | The rate (30–50 Hz) follows how firmly you press the impulse head. Consider it for larger muscle groups and extremities. Force up to 20. |
| Point | Sweeps through the rates from 4 to 90 Hz and back. Consider it for trigger points, and for patients who are not obese. Force up to 20. |
On the impulse head, double-click the button to switch between Osseous, Point and Myofascial, and press and hold it to step up the force. PulStar announces force and mode changes aloud. The force, mode and rate can't be changed while impulses are being delivered.
The impulses also stop on their own when:
Myofascial mode doesn't use Max Number Of Impulses: it continues until you lift the impulse head, up to 999 impulses.
The screen shows the impulses delivered to the current vertebra and the total for the session.
PulStar doesn't move on by itself; you decide when each vertebra is done. You can treat a vertebra as many times as you need. When you're ready:
After the last vertebra, the next click of the button moves to the post-analysis.
The impulse head is designed for 2,000 impulses within 5 minutes, followed by 2 minutes resting in its holder, where a fan cools it. PulStar changes the color of the impulse count for about 2 minutes after every 2,000 impulses as a reminder. If the back of the impulse head is too hot to hold comfortably, put it in the holder until it cools; overheating can damage the force sensor.
Soft tissue uses its own default force. Edit the list of locations in Advanced preferences. Don't use PulStar on the soft tissues listed in Safety and contraindications.