Repeatable results depend on a working instrument. Once a month, test it on your own hand:
Your results should look similar to this example. The bars step down from the heel of the hand, to the palm, to the fingertip, and the bars for each spot are close to each other.
The exact values don't matter, since every hand is different. If your results don't resemble the example, the instrument may need recalibration or the sensor may be damaged; call Sense Technology at +1 724-733-2277, with the serial number ready.
The only service you can do yourself is checking and replacing the two fuses in the power entry module on the back of the PulStar:
All other maintenance, calibration and repair must be done by a Sense Technology certified PulStar repair center. Have the system serviced every 3 years. Call Sense Technology at +1 724-733-2277 to find the nearest repair center.
| Temperature | 50–95 °F (10–35 °C) |
| Relative humidity | 20–80%, non-condensing |
| Altitude | 0–10,000 ft (0–3,048 m) |
| Power | 115/230 V AC, 50/60 Hz, single phase |
Dispose of the PulStar according to local laws. When it reaches the end of its life, contact Sense Technology or your local authorities about recycling. In the European Union, take it to a designated collection point rather than disposing of it with household waste.